top of page
IOM Nihon Kohden


MEE-2000
-
ABR 17HZ/80Db stimulerer trommehinne
-
Signaler fra trommehinne til cochlearnervens inngang til hjernestammen
-
Fanges også opp av DNAP-elekroden ved hjernestammen og averages
-
FRENAP-elektrode stimulerer facialnervens utgang fra hjernestammen. 1-3mA 0,1HZ
-
EMG av oris og oculi
-
18 bit A/D conversion
-
Up to 4 breakout boxes connection
-
Able to disconnect box during acquisition
-
32 inputs under the cover for direct inputs
-
SpO2 and EtCO2 connection (Option)
-
In-built stimulator on main unit
-
Constant current & Constant voltage
-
Daisy chain stimulator up to 4 pods
-
Each pod has 8 pairs for High and 2 for Low
-
TcMEP Matrix pod (Option) or BM-122B
-
for TcMEP stimulation
-
All 8 inputs can be programmed for anodal or cathodal
-
(JS-201B)
-
Constant current : 0 to 250mA (0.2mA steps)
-
Constant voltage : 0 to 1,000V (1V steps)
-
Pulse Duration : 0.05, 0.1, 0.2, 0.3, 0.5, 0.7, 1ms
-
Polarity : Positive, Negative
-
Phase : Monophasic, Biphasic, Alternating


Spill video